Ontology highlight
ABSTRACT: Purpose
To compare aqueous humour (AH) dynamics in the presence of a precrystalline (Implantable Collamer Lens®; ICL) or iris-fixed (Artiflex®) phakic intraocular lens (PIOL).Methods
By computational fluid dynamics simulation, AH flow was modelled through a peripheral iridotomy (PI) or central lens hole (both 360 μm) in the presence of an Artiflex or ICL lens, respectively. The impacts of AH flow were then determined in terms of wall shear stress (WSS) produced on the endothelium or crystalline lens. Effects were also modelled for different scenarios of pupil diameter (PD 3.5 or 5.5 mm), ICL vault (100, 350, 800 μm) and number of Artiflex iridotomies (1 or 2) and location (12 or 6 o'clock).Results
For a PD of 3.5 mm, AH volumes flowing from the posterior to the ante
SUBMITTER: Fernandez-Vigo JI
PROVIDER: S-EPMC6089426 | biostudies-literature | 2018
REPOSITORIES: biostudies-literature