Arterial compliance probe for cuffless evaluation of carotid pulse pressure.


ABSTRACT:

Objective

Assessment of local arterial properties has become increasingly important in cardiovascular research as well as in clinical domains. Vascular wall stiffness indices are related to local pulse pressure (ΔP) level, mechanical and geometrical characteristics of the arterial vessel. Non-invasive evaluation of local ΔP from the central arteries (aorta and carotid) is not straightforward in a non-specialist clinical setting. In this work, we present a method and system for real-time and beat-by-beat evaluation of local ΔP from superficial arteries-a non-invasive, cuffless and calibration-free technique.

Methods

The proposed technique uses a bi-modal arterial compliance probe which consisted of two identical magnetic plethysmograph (MPG) sensors located at 23 mm distance