Unknown

Dataset Information

Practical application of Octavius® -4D: Characteristics and criticalities for IMRT and VMAT verification.


ABSTRACT: Octavius® -4D is a very effective device in radiotherapy treatment quality assurance (QA), due to its simple set-up and analysis package. However, even if it is widely used, its main characteristics and criticalities were only partially investigated. Taking start from its commissioning, the aim of this work was to study the main dependencies of the device response. The outcome dependence was studied comparing results by different delivery techniques [Intensity Modulated Radiation Therapy, IMRT (n = 29) and RapidArc, RA (n = 15)], anatomical regions [15 head/neck, 19 pelvis and 10 pancreas] and linear accelerators [DHX (n = 14) and Trilogy (n = 30)]. Moreover, the agreement dependency on the section of the phantom was assessed. Plan evaluations obtained by 2D, 3D, and volumetric

SUBMITTER: Urso P 

PROVIDER: S-EPMC6123102 | biostudies-literature | 2018 Sep

REPOSITORIES: biostudies-literature