Fracture of an S-ROM stem at the sleeve-stem junction.


ABSTRACT: Fracture of a well-ingrown femoral component is a rare and often challenging complication. Modular junctions and sleeve interfaces have been identified as one potential point of weakness with corrosion and fretting being contributing factors to ultimate femoral component fracture. Stem fractures at the sleeve interface were reported occasionally for the proximal ingrowth modular Emperion System (Smith and Nephew, Memphis, TN). However, this failure mechanism has been reported infrequently, often associated with corrosion at the modular junction, for the similarly designed S-ROM system (DePuy Orthopedics Inc., Warsaw, IN). We present the case of a 52-year-old patient, with a body weight of 84 kg (185 lbs) and a body mass index of 30.6 kg/m2, who suffered a fatigue fracture of a 1

SUBMITTER: Rueckl K 

PROVIDER: S-EPMC6123181 | biostudies-literature | 2018 Sep