Interdevice variability of central corneal thickness measurement.


ABSTRACT:

Purpose

To evaluate variability of central corneal thickness measurement (CCT) devices using a hitherto unprecedented number of CCT devices.

Methods

CCT was measured consecutively in 122 normal corneas of 61 subjects with seven different devices using three distinct measurement technologies: Scheimpflug, Ultrasound, and Optical Coherence Tomography (OCT). Per device deviation from the mean CCT value per eye was used to determine which of the devices performed best, compared to the mean value.

Results

Cirrus OCT yielded the lowest deviation. Deviations of the individual devices from the mean CCT of each eye were (OS/OD) 12.8±5.0/14.9±9.4 μm for Topcon noncontact specular microscopy (NCSM), 11.3±5.9/10.6±7.3 μm for Pentacam, 10.7±5.2/10.4±4.8 μm for Spectralis OCT, 6.0±