Noninvasive Blood Pressure Monitor Designed for Patients With Heart Failure Supported with Continuous-Flow Left Ventricular Assist Devices.
Ontology highlight
ABSTRACT: The gold standard for noninvasive blood pressure (BP) measurement, the Doppler technique, does not provide systolic blood pressure (SBP) and diastolic blood pressure (DBP) and may limit therapy outcomes. To improve patient care, we tested specifically designed experimental BP (ExpBP) monitor and the Doppler technique by comparing noninvasive measures to the intraarterial (I-A) BP in 31 patients with end-stage heart failure (4 females) 2.6 ± 3.4 days post-LVAD implantation (20 HeartMate II and 11 HeartWare). Bland-Altman plots revealed that the ExpBP monitor overestimated mean arterial pressure (MAP) by 1.2 (4.8) mm Hg (mean difference [standard deviation]), whereas the Doppler by 6.7 (5.8) mm Hg. The ExpBP SBP was overestimated by 0.8 (6.1) mm Hg and DBP by 1.9 (5.3) mm Hg compared with th