Ontology highlight
ABSTRACT: Background
The process of assuring the safety of medical devices is constrained by reliance on voluntary reporting of adverse events. We evaluated a strategy of prospective, active surveillance of a national clinical registry to monitor the safety of an implantable vascular-closure device that had a suspected association with increased adverse events after percutaneous coronary intervention (PCI).Methods
We used an integrated clinical-data surveillance system to conduct a prospective, propensity-matched analysis of the safety of the Mynx vascular-closure device, as compared with alternative approved vascular-closure devices, with data from the CathPCI Registry of the National Cardiovascular Data Registry. The primary outcome was any vascular complication, which was a compos
SUBMITTER: Resnic FS
PROVIDER: S-EPMC6145132 | biostudies-literature | 2017 Feb
REPOSITORIES: biostudies-literature