Measurable Range of Subfoveal Choroidal Thickness With Conventional Spectral Domain Optical Coherence Tomography.


ABSTRACT:

Purpose

We investigate the measurable range of subfoveal choroidal thickness (SFCT) with conventional spectral domain optical coherence tomography (SD-OCT) in normal healthy eyes.

Methods

All subjects underwent an SD-OCT horizontal line scan centered at the fovea with standard and enhanced depth imaging (EDI) techniques. Two independent observers manually measured SFCT on standard and EDI-OCT images, if two choroidal borders were identified. The rate of successful measurement with standard OCT was evaluated. Inter- and intraclass correlation coefficients and Bland-Altman plots were used for analysis.

Results

OCT images of 724 normal healthy eyes from 374 subjects were evaluated. Although the inner choroidal border was identified on all OCT images, the outer choroidal