ABSTRACT: Background
Clinicians in intensive care units experience alarm fatigue related to frequent false and non-actionable alarms produced by physiologic monitors. To reduce non-actionable alarms, alarm settings may need to be customized for individual patients; however, nurses may not customize alarms because of competing demands and alarm fatigue.Objective
To examine the effectiveness and acceptance of physiologic monitor software to support customization of alarms.Methods
This pre/post intervention study was conducted in a 56-bed medical intensive care unit. IntelliVue® Alarm Advisor customization support software for alarm limit violations was installed on all monitors and education on its use provided. For 2 months before and after implementation of the software, data