ABSTRACT: Purpose
To evaluate the repeatability, the reproducibility and the agreement of foveal avascular zone (FAZ) measurements using three different optical coherence tomography angiography (OCT-A) devices.Procedures
This prospective study included 24 eyes of 24 healthy volunteers. OCT-A imaging was performed using RTVue XR Avanti, Canon OCT-HS100 and Spectralis OCT-A. Repeated measurements were performed under the same conditions on two separate days, and the area of the FAZ was determined and analyzed using the above devices.Results
All three devices showed a high ICC and there was no significant difference between the ICCs (pairwise comparison) of the three devices (Optovue-Canon (p = 0.66); Canon-Heidelberg (p = 0.21); Heidelberg-Optovue (p = 0.37). Agreement analysis