Assessment of the performance of blood glucose monitoring systems for monitoring dysglycaemia in neonatal patients.


ABSTRACT:

Objective

To validate a three-step protocol that assesses the clinical risk associated with using blood glucose monitoring systems (BGMS) in neonates for the management of dysglycaemia.

Method

The three-step validation approach included confirmation of the accuracy of the reference method using National Institute of Standards and Technology (NIST) glucose standards, assessment of analytical risk performed on whole blood collected from paediatric patients routinely tested for glucose and a clinical risk assessment performed using heel stick capillary samples collected from 147 new-born babies and neonates admitted to intensive care. BGMS glucose measurements were compared with the NIST aligned laboratory reference method.

Results

The accuracy of the laboratory referenc