Abdominal Positioning of the Next-Generation Intra-Aortic Fluid Entrainment Pump (Aortix) Improves Cardiac Output in a Swine Model of Heart Failure.


ABSTRACT:

Background

Acute heart failure refractory to medical therapy is a major cause of morbidity and mortality. The Aortix device (Procyrion Inc) is a percutaneously delivered entrainment pump positioned in the descending aorta.

Methods and results

Using the newest generation Aortix device in 8 adult male Yorkshire swine, we tested the hypothesis that positioning in the abdominal aorta may provide superior hemodynamic effects than thoracic positioning in a swine model of postinfarct left ventricular injury.Abdominal activation generated significantly larger transaortic gradients (proximal minus distal mean aortic pressures) than thoracic positioning at all pump speeds. Compared with baseline values, activation in the abdominal, not thoracic, position significantly increased cardia