Ontology highlight
ABSTRACT: Aim
To determine the influence of the construction design over the biological component's performance in an experimental bio-artificial liver (BAL) device.Methods
Two BAL models for liver microorgans (LMOs) were constructed. First, we constructed a cylindrical BAL and tested it without the biological component to establish its correct functioning. Samples of blood and biological compartment (BC) fluid were taken after 0, 60, and 120 min of perfusion. Osmolality, hematocrit, ammonia and glucose concentrations, lactate dehydrogenase (LDH) release (as a LMO viability parameter), and oxygen consumption and ammonia metabolizing capacity (as LMO functionality parameters) were determined. CPSI and OTC gene expression and function were measured. The second BAL, a "flat bottom" mode
SUBMITTER: Pizarro MD
PROVIDER: S-EPMC6206149 | biostudies-literature | 2018 Oct
REPOSITORIES: biostudies-literature