Ontology highlight
ABSTRACT: Background
A novel multiparameter brain sensor (MPBS) allows the simultaneous measurement of brain tissue oxygenation (ptiO2), cerebral blood flow (CBF), intracranial pressure (ICP), and brain temperature with a single catheter. This laboratory investigation evaluates the MPBS in an animal model in relation to established reference probes.Methods
The study group consisted of 17 juvenile male pigs. Four MPBS and four reference probes were implanted per pig and compared simultaneously. The measured parameters were challenged by standardized provocations such as hyperoxia, dobutamine, and norepinephrine application, hypercapnia and hypoxia in combination with and without a controlled cortical impact (CCI) injury. Mean values over 2 min were collected for predefined
SUBMITTER: Mader MM
PROVIDER: S-EPMC6208836 | biostudies-literature | 2018 Oct
REPOSITORIES: biostudies-literature