Reliability of Compartmental Body Composition Measures in Weight-Stable Adults Using GE iDXA: Implications for Research and Practice.
Ontology highlight
ABSTRACT: The aim of this study was to explore the reliability and precision of body compartment measures, in particular visceral adipose tissue, in weight stable adults over a range of BMIs using GE-Lunar iDXA. Weight-stable participants aged 18⁻65 years had a total body composition scan on GE-Lunar iDXA either on three separate occasions over a three month period (n = 51), or on a single occasion for duplicate scans with repositioning (n = 30). The coefficient of variation (CV%) and least significant change (LSC) of body compartments were calculated. The CV was higher for all measures over three months (range 0.8⁻5.9%) compared with same-day precision-scans (all < 2%). The CV for visceral adipose tissue (VAT) was considerably higher than all other body compartments (42.2% three months, 16.2% same
SUBMITTER: Dordevic AL
PROVIDER: S-EPMC6213248 | biostudies-literature | 2018 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA