Comparison of hemodialysis arteriovenous fistula blood flow rates measured by Doppler ultrasound and phase-contrast magnetic resonance imaging.


ABSTRACT:

Objective

The objective of this study was to compare blood flow rates measured by Doppler ultrasound (DUS) and phase-contrast magnetic resonance imaging (MRI) in patients having a hemodialysis arteriovenous fistula (AVF) and to identify scenarios in which there was significant discordance between these two approaches.

Methods

Blood flow rates in the proximal artery (PA) and draining vein (DV) of newly created upper extremity AVFs were measured and compared using DUS and phase-contrast MRI at 1 day, 6 weeks, and 6 months postoperatively.

Results

Blood flow rates in the PA measured by DUS (1155 ± 907 mL/min, mean ± standard deviation) and by MRI (1170 ± 657 mL/min) were not statistically different (P = .812) based on 78 data pairs from 49 patients. DV DUS flow (1277 ± 9