ABSTRACT: Purpose
To compare the performance of the pattern standard deviation (PSD) values derived from the central 12 locations of the 24-2 visual field test (C24-2) to the entire 10-2 test for detecting central visual field abnormalities in eyes with, suspected of having, or at risk of having glaucoma.Design
Cross-sectional case-control study.Methods
Eyes with, suspected of having, or at risk of having glaucoma, based on masked grading of optic disc stereophotographs and/or ocular hypertension (intraocular pressure ≥ 22 mm Hg) were included as cases (n = 523). Eyes from healthy participants were included as controls (n = 107) to allow the 2 tests to be compared at matched specificities. The sensitivity to detect cases at 95% specificity using PSD values derived from the en