Design and Validation of a Semi-Active Variable Stiffness Foot Prosthesis.


ABSTRACT: This paper presents the design and validation of a novel lower limb prosthesis called the variable stiffness foot (VSF), designed to vary its forefoot stiffness in response to user activity. The VSF is designed as a semi-active device that adjusts its stiffness once per stride during swing phases, in order to minimize size, mass, and power consumption. The forefoot keel is designed as an overhung composite beam, whose stiffness is varied by moving a support fulcrum to change the length of the overhang. Stiffness modulation is programmed in response to the gait characteristics detected through foot trajectory reconstruction based on an embedded inertial sensor. The prototype VSF has a mass of only 649 g including the battery, and a build height of 87 mm. Mechanical testing demonstrated a fo