Performance evaluation of RF coils integrated with an RF-penetrable PET insert for simultaneous PET/MRI.


ABSTRACT:

Purpose

An "RF-penetrable" PET insert that allows the MR body coil to be used for RF transmission was developed to make it easier for an existing MR center to achieve simultaneous PET/MRI. This study focuses on experiments and analyses to study PET/RF coil configurations for simultaneous PET/MR studies.

Methods

To investigate the appropriate RF coil design, a transmit/receive (TX/RX) birdcage coil and an RX-only phased-array coil (TX from body coil), both fitting inside the PET ring were built and characterized. For MR performance evaluation, B1 field uniformity and MR image SNR were calculated. PET photon attenuation due to each coil was studied by means of CT-based attenuation maps and reconstructed PET images.

Results

When using the RX-only phased-array

SUBMITTER: Lee BJ 

PROVIDER: S-EPMC6289759 | biostudies-literature | 2019 Feb

REPOSITORIES: biostudies-literature