ABSTRACT: Spiroergometric measurements of persons who require oxygen insufflation due to illness can be performed under conditions of increased inspiratory oxygen concentration (FIO2). This increase in FIO2, however, often leads to errors in the calculation of oxygen consumption ([Formula: see text]). These inconsistencies are due to the application of the Haldane Transformation (HT), an otherwise indispensable correction factor in the calculation of [Formula: see text] that becomes inaccurate at higher FIO2 concentrations. A possible solution to this problem could be the use of the 'Eschenbacher transformation' (ET) as an alternative correction factor. This study examines the concentration of FIO2 at which the HT and the ET are valid, providing plausible data of oxygen consumption corresponding to