ABSTRACT: Aim
To evaluate Attain Performa (Medtronic, Dublin, Ireland) quadripolar lead performance in clinical practice and, secondarily, to compare its long term clinical outcomes vs bipolar leads for left ventricular (LV) pacing.Methods and results
We retrospectively analyzed clinical, procedural and follow-up data of 215 patients implanted with a quadripolar lead. One hundred and twenty one patients implanted with bipolar lead were selected to compare long-term clinical outcomes. The quadripolar lead was implanted in the target vein in 196 patients (91%) without acute dislodgements. In 50% of patients the chosen final pacing configuration at implant would not have been available with bipolar leads. A dedicated quadripolar pacing vector was chosen more frequently when the LV tip l