Impact of advanced detector technology and iterative reconstruction on low-dose quantitative assessment of lung computed tomography density in a biological lung model.
Ontology highlight
ABSTRACT: Quantitative computed tomography (QCT)-derived measures of lung density are valued methods for objectively characterizing lung parenchymal and peripheral airways disease and are being used in a growing number of lung disease focused trials. Detector and reconstruction improvements in CT technology have allowed for significant radiation dose reduction in image acquisition with comparable qualitative image quality. We report the impact of detector type and reconstruction type on QCT lung density measures in relation to decreasing dose indices. Two sets of studies were completed in an in vivo pig model with a SOMATOM Definition Flash CT system: (a) prior to system upgrade with conventional detectors (UFC) and filtered back projection (FBP), and (b) post system upgrade with integrated electron
SUBMITTER: Hammond E
PROVIDER: S-EPMC6309498 | biostudies-literature | 2018 Jun