Interlaboratory Reproducibility of a Targeted Metabolomics Platform for Analysis of Human Serum and Plasma.


ABSTRACT: A critical question facing the field of metabolomics is whether data obtained from different centers can be effectively compared and combined. An important aspect of this is the interlaboratory precision (reproducibility) of the analytical protocols used. We analyzed human samples in six laboratories using different instrumentation but a common protocol (the AbsoluteIDQ p180 kit) for the measurement of 189 metabolites via liquid chromatography (LC) or flow injection analysis (FIA) coupled to tandem mass spectrometry (MS/MS). In spiked quality control (QC) samples 82% of metabolite measurements had an interlaboratory precision of <20%, while 83% of averaged individual laboratory measurements were accurate to within 20%. For 20 typical biological samples (serum and plasma from healthy indivi