Assessing the accuracy of commercially available gas sensors for the measurement of ambient ozone and nitrogen dioxide.
Ontology highlight
ABSTRACT: The objective of the National Institute for Occupational Safety and Health (NIOSH) accuracy criterion is to ensure that measurements from monitoring devices are within ±25% of the true concentration of the analyte with 95% certainty. To determine whether NO2 and O3 sensors meet this criterion, three commercially available units (Cairclip O3/NO2, Aeroqual NO2, and Aeroqual O3 sensors) were co-located three times with validated instruments (NOx chemiluminescence [NO2mon] and photometric O3 analyzers [O3mon]) at an outdoor monitoring station. As cofactors of sensor performance such as temperature (T) and relative humidity (RH) potentially influence the response of NO2 and O