Ontology highlight
ABSTRACT: Purpose
Abdominal MRI remains challenging because of respiratory motion. Motion compensation strategies are difficult to compare clinically because of the variability across human subjects. The goal of this study was to evaluate a programmable system for one-dimensional motion management MRI research.Methods
A system comprised of a programmable motorized linear stage and computer was assembled and tested in the MRI environment. Tests of the mutual interference between the platform and a whole-body MRI were performed. Organ trajectories generated from a high-temporal resolution scan of a healthy volunteer were used in phantom tests to evaluate the effects of motion on image quality and quantitative MRI measurements.Results
No interference between the motion platform
SUBMITTER: Nofiele J
PROVIDER: S-EPMC6342555 | biostudies-literature | 2016 Aug
REPOSITORIES: biostudies-literature