High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers.
Ontology highlight
ABSTRACT: PURPOSE:Objective of this study was to evaluate the effectiveness of the iterative reconstruction of high-pitch dual-source chest CT (IR-HP-CT) scanned with low radiation exposure compared with low dose chest CT (LDCT). MATERIALS AND METHODS:This study was approved by the institutional review board. Thirty healthy adult volunteers (mean age 44 years) were enrolled in this study. All volunteers underwent both IR-HP-CT and LDCT. IR-HP-CT was scanned with 120 kVp tube voltage, 30 mAs tube current and pitch 3.2 and reconstructed with sinogram affirmed iterative reconstruction. LDCT was scanned with 120 kVp tube voltage, 40 mAs tube current and pitch 0.8 and reconstructed with B50 filtered back projection. Image noise, and signal to noise ratio (SNR) of the infraspinatus muscle, subcutaneous fa