Optimal threshold of three-dimensional echocardiographic fully automated software for quantification of left ventricular volumes and ejection fraction: Comparison with cardiac magnetic resonance disk-area summation method and feature tracking method.


ABSTRACT:

Aims

Novel fully automated left chamber quantification software for three-dimensional echocardiography (3DE) has a potential for reliable measurement of left ventricular (LV) volumes and ejection fraction (LVEF). However, the optimal setting of global LV endocardial border threshold has not been settled.

Methods and results

We performed LV volumes and LVEF analysis using fully automated left chamber quantification software (Dynamic HeartModelA.I., Philips Medical Systems) in 65 patients who had undergone both 3DE and cardiac magnetic resonance (CMR) examinations on the same day. We recorded LV end-diastolic volume (LVEDV) and LV end-systolic volume (LVESV) according to the change in LV global border threshold settings from 0-point to 100-point with each increment of 10-point

SUBMITTER: Wu VC 

PROVIDER: S-EPMC6349335 | biostudies-literature | 2019