Analyses of the factors influencing the accuracy of three-dimensional ultrasound in comparison with cone-beam CT in image-guided radiotherapy for prostate cancer with or without pelvic lymph node irradiation.


ABSTRACT:

Background

Three-dimensional ultrasound (3DUS) is an attractive option in image-guided radiotherapy (IGRT) for prostate cancer (PCa) patients. However, the potential factors influencing the accuracy of 3DUS in comparison with cone-beam CT (CBCT) in IGRT for PCa patients haven't been clearly identified.

Methods

The differences between US/US and CBCT/CT registrations were analyzed over 586 and 580 sessions for 24 and 25 PCa patients treated with or without pelvic lymph node irradiation, respectively. The clinical factors that may influence registration differences were also evaluated.

Results

The average discrepancies between US/US and CBCT/CT registrations were - 0.28 ± 5.28 mm, - 0.16 ± 3.48 mm, and - 0.47 ± 4.31 mm in the superior-inferior (SI), left-right (LR), and