Free-breathing, non-ECG, continuous myocardial T1 mapping with cardiovascular magnetic resonance multitasking.


ABSTRACT:

Purpose

To evaluate the accuracy and repeatability of a free-breathing, non-electrocardiogram (ECG), continuous myocardial T1 and extracellular volume (ECV) mapping technique adapted from the Multitasking framework.

Methods

The Multitasking framework is adapted to quantify both myocardial native T1 and ECV with a free-breathing, non-ECG, continuous acquisition T1 mapping method. We acquire interleaved high-spatial resolution image data and high-temporal resolution auxiliary data following inversion-recovery pulses at set intervals and perform low-rank tensor imaging to reconstruct images at 344 inversion times, 20 cardiac phases, and 6 respiratory phases. The accuracy and repeatability of Multitasking T1 mapping in generating nativ