Nanoscale Ion Emitters in Native Mass Spectrometry for Measuring Ligand-Protein Binding Affinities.
Ontology highlight
ABSTRACT: Electrospray ionization (ESI) mass spectrometry (MS) is a crucial method for rapidly determining the interactions between small molecules and proteins with ultrahigh sensitivity. However, nonvolatile molecules and salts that are often necessary to stabilize the native structures of protein-ligand complexes can readily adduct to protein ions, broaden spectral peaks, and lower signal-to-noise ratios in native MS. ESI emitters with narrow tip diameters (∼250 nm) were used to significantly reduce the extent of adduction of salt and nonvolatile molecules to protein complexes to more accurately measure ligand-protein binding constants than by use of conventional larger-bore emitters under these conditions. As a result of decreased salt adduction, peaks corresponding to protein-ligand complexes t