Unknown

Dataset Information

Parameter map error due to normal noise and aliasing artifacts in MR fingerprinting.


ABSTRACT:

Purpose

To introduce a quantitative tool that enables rapid forecasting of T1 and T2 parameter map errors due to normal and aliasing noise as a function of the MR fingerprinting (MRF) sequence, which can be used in sequence optimization.

Theory and methods

The variances of normal noise and aliasing artifacts in the collected signal are related to the variances in T1 and T2 maps through derived quality factors. This analytical result is tested against the results of a Monte-Carlo approach for analyzing MRF sequence encoding capability in the presence of aliasing noise, and verified with phantom experiments at 3 T. To further show the utility of our approach, our quality factors are used to find efficient MRF sequences for fewer repe

SUBMITTER: Kara D 

PROVIDER: S-EPMC6414267 | biostudies-literature | 2019 May

REPOSITORIES: biostudies-literature