Noninvasive Self-diagnostic Device for Tear Collection and Glucose Measurement.


ABSTRACT: We propose a noninvasive, self-diagnostic device that enables safe tear collection and glucose measurement. The device described herein was manufactured by tight assembly of a lid for tear collection in conjunction with a strip-type glucose sensor. The lid was designed to be in contact with the inferior palpebral conjunctiva for tear collection and was thus designed to possess a proper contact area and rounded boundaries to avoid eye tissue damage. For the strip-type glucose sensor, we employed a commercially available electrochemical sensor (Accu-Chek test strips), which was modified to reduce the volume of the reaction chamber (0.4 μl) for a small amount of collected tear fluid. When tested with in vivo animal models, the device was able to collect tear fluid in a relatively short time (