Reduction of Metal Artifacts and Improvement in Dose Efficiency Using Photon-Counting Detector Computed Tomography and Tin Filtration.


ABSTRACT: OBJECTIVES:The aim of this study was to investigate the impact on metal artifacts and dose efficiency of using a tin filter in combination with high-energy threshold (TH) images of a photon-counting detector (PCD) computed tomography (CT) system. MATERIALS AND METHODS:A 3D-printed spine with pedicle screws was scanned on a PCD-CT system with and without tin filtration. Image noise and severity of artifacts were measured for low-energy threshold (TL) and TH images. In a prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant study, 20 patients having a clinical energy-integrating detector (EID) CT were scanned on a PCD-CT system using tin filtration. Images were reviewed by 3 radiologists to evaluate visualization of anatomic structure