The added value of three-dimensional echocardiography in the late diagnosis of a pacemaker complication in a patient with severe congestive heart failure: a case report.


ABSTRACT:

Background

Three-dimensional echocardiography (3DE) presents an increasingly important role in the management of interventional cardiac procedures, overcoming limitations of conventional two-dimensional echocardiography (2DE). Early use of 3DE might have an added value in the diagnosis of device-related complications, such as lead induced tricuspid regurgitation (LITR), by providing better understanding of its mechanisms and ensuring a prompt and individually tailored treatment strategy.

Case summary

We report the case of a female patient with repeated hospitalizations for congestive heart failure in the past 2 years, who had a permanent single-chamber ventricular pacemaker (PM) implanted 10 years ago and a misleading diagnosis of severe tricuspid regurgitation (TR) secondar