Ontology highlight
ABSTRACT: Objective
The MiniMed 670G System is the first commercial hybrid closed-loop (HCL) system for management of type 1 diabetes. Using data from adolescent and young adult participants, we compared insulin delivery patterns and time-in-range metrics in HCL (Auto Mode) and open loop (OL). System alerts, usage profiles, and operational parameters were examined to provide suggestions for optimal clinical use of the system.Research design and methods
Data from 31 adolescent and young adult participants (14-26 years old) at three clinical sites in the 670G pivotal trial were analyzed. Participants had a 2-week run-in period in OL, followed by a 3-month in-home study phase with HCL functionality enabled. Data were compared between baseline OL and HCL use after 1 week, 1 month, 2 mont
SUBMITTER: Messer LH
PROVIDER: S-EPMC6463622 | biostudies-literature | 2018 Apr
REPOSITORIES: biostudies-literature