Response Surface Methodology for the Optimisation of Electrochemical Biosensors for Heavy Metals Detection.
Ontology highlight
ABSTRACT: Herein, we report the application of a chemometric tool for the optimisation of electrochemical biosensor performances. The experimental design was performed based on the responses of an amperometric biosensor developed for metal ions detection using the flow injection analysis. The electrode preparation and the working conditions were selected as experimental parameters, and thus, were modelled by a response surface methodology (RSM). In particular, enzyme concentration, flow rates, and number of cycles were reported as continuous factors, while the sensitivities of the biosensor (S, µA·mM-1) towards metals, such as Bi3+ and Al3+ were collected as responses and optimised by a central composite design (CCD). Bi3+ and Al3+ inhibition on the Pt/PPD/GOx biosensor response is for the first tim
SUBMITTER: De Benedetto GE
PROVIDER: S-EPMC6468913 | biostudies-literature | 2019 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA