ABSTRACT: Objective
The study aims at evaluating mid-infrared human milk analyzer (HMA) accuracy and precision, in human milk (HM).Study design
Röse-Gottlieb, high-performance anion exchange chromatography-pulsed amperometric detection (HPAEC-PAD), Kjeldahl and amino acid analysis (AA) were selected as references for total fat, lactose and total protein determination.Results
No significant difference was observed in lactose content between HMA and HPAEC-PAD. Significant differences were observed in fat and protein content between HMA and reference methods. However, the difference in fat content was lower than 12%, and therefore within the variability declared by supplier. For protein determination, the BCA protein assay was selected. No significant differences were observed i