Electron streams in air during magnetic-resonance image-guided radiation therapy.
Ontology highlight
ABSTRACT: To investigate the undesired irradiations outside of the treatment field by electron streams in air (air-electron-stream) during magnetic-resonance image-guided radiation therapy (MR-IGRT). A custom-made support phantom adjusting angles between the beam central axis (CAX) and the phantom surface (termed phantom-angles), were used. Using the ViewRay system, a rectangular parallelepiped phantom placed on the support phantom, was irradiated with field sizes of 6.3 cm × 6.3 cm (FS6.3) and 12.6 cm × 12.6 cm (FS12.6) at gantry angles of 0°, 30°, and 330°, and phantom-angles of 10°, 20°, and 30°. For each beam delivery, the isocenter was located at the center of mass of the phantom and 3 Gy was delivered to the isocenter (prescription dose = 3 Gy). The doses given by the air-electron-streams were