Evaluating Separation Selectivity and Collision Cross Section Measurement Reproducibility in Helium, Nitrogen, Argon, and Carbon Dioxide Drift Gases for Drift Tube Ion Mobility-Mass Spectrometry.


ABSTRACT: Previous ion mobility (IM) studies have demonstrated that varying the drift gas composition can be used to enhance chemical selectivity and resolution, yet there are few drift gas studies aimed at achieving quantitatively reproducible mobility measurements. Here, we critically evaluate the conditions necessary to achieve reproducible collision cross section (CCS) measurements in pure drift gases (helium, nitrogen, argon, and carbon dioxide) using a commercial uniform field drift tube instrument. Optimal experimental parameters are assessed based on the convergence of CCS measurements to reproducible values which are compared with literature values. A suite of calibration standards with diverse masses, biological classes, and charge states are examined to assess chemical selectivity and res

SUBMITTER: Morris CB 

PROVIDER: S-EPMC6520154 | biostudies-literature | 2019 Jun

REPOSITORIES: biostudies-literature