ABSTRACT: Objective
The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM).Methods
A total of 333 blood samples from different subgroups (168 cases with thrombocytopenia, 136 cases with normal platelet counts and 29 cases with thrombocytosis) were tested. Regarding IRM as the gold standard, we compared the accuracy and precision of different platelet count methods; i.e. LH780 (impedance), BC-6000 Plus (optical (O) and impedance (I)), Sysmex XN-9000 (optical (O), impedance (I), fluorescence (F)), and hand counting.Results
Sysmex XN-9000-F (r = 0.988) had the best correlation with IRM