Cardiac Rhythm Device Identification Using Neural Networks.


ABSTRACT:

Objectives

This paper reports the development, validation, and public availability of a new neural network-based system which attempts to identify the manufacturer and even the model group of a pacemaker or defibrillator from a chest radiograph.

Background

Medical staff often need to determine the model of a pacemaker or defibrillator (cardiac rhythm device) quickly and accurately. Current approaches involve comparing a device's radiographic appearance with a manual flow chart.

Methods

In this study, radiographic images of 1,676 devices, comprising 45 models from 5 manufacturers were extracted. A convolutional neural network was developed to classify the images, using a training set of 1,451 images. The testing set contained an additional 225 images consisting of 5 examples of each model. The network's ability to identify the manufacturer of a device was compared with that of cardiologists, using a published flowchart.

Results

The neural network was 99.6% (95% confidence interval [CI]: 97.5% to 100.0%) accurate in identifying the manufacturer of a device from a radiograph and 96.4% (95% CI: 93.1% to 98.5%) accurate in identifying the model group. Among 5 cardiologists who used the flowchart, median identification of manufacturer accuracy was 72.0% (range 62.2% to 88.9%), and model group identification was not possible. The network's ability to identify the manufacturer of the devices was significantly superior to that of all the cardiologists (p < 0.0001 compared with the median human identification; p < 0.0001 compared with the best human identification).

Conclusions

A neural network can accurately identify the manufacturer and even model group of a cardiac rhythm device from a radiograph and exceeds human performance. This system may speed up the diagnosis and treatment of patients with cardiac rhythm devices, and it is publicly accessible online.

SUBMITTER: Howard JP 

PROVIDER: S-EPMC6537849 | biostudies-literature | 2019 May

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Objectives</h4>This paper reports the development, validation, and public availability of a new neural network-based system which attempts to identify the manufacturer and even the model group of a pacemaker or defibrillator from a chest radiograph.<h4>Background</h4>Medical staff often need to determine the model of a pacemaker or defibrillator (cardiac rhythm device) quickly and accurately. Current approaches involve comparing a device's radiographic appearance with a manual flow chart.<h4  ...[more]

Similar Datasets

| S-EPMC8850959 | biostudies-literature
| S-EPMC5603510 | biostudies-literature
| S-EPMC9448979 | biostudies-literature
| S-EPMC8408627 | biostudies-literature
| S-EPMC4921197 | biostudies-literature
| S-EPMC8765108 | biostudies-literature
| S-EPMC10630086 | biostudies-literature
| S-EPMC8718715 | biostudies-literature
| S-EPMC7531966 | biostudies-literature
| S-EPMC5751783 | biostudies-literature