ABSTRACT: Background
In electric-acoustic pitch matching experiments in patients with single-sided deafness and a cochlear implant, the observed "mismatch" between perceived pitch and predicted pitch, based on the amended Greenwood frequency map, ranges from -1 to -2 octaves. It is unknown if and how this mismatch differs for perimodiolar versus lateral wall electrode arrays.Objectives
We aimed to investigate if the type of electrode array design is of influence on the electric-acoustic pitch match.Method
Fourteen patients (n = 8 with CI422 + lateral wall electrode array, n = 6 with CI512 + perimodiolar electrode array; Cochlear Ltd.) compared the pitch of acoustic stimuli to the pitch of electric stimuli at two test sessions (average interval 4.3 months). We plotted these "p