Lag Time Remains with Newer Real-Time Continuous Glucose Monitoring Technology During Aerobic Exercise in Adults Living with Type 1 Diabetes.


ABSTRACT: Background: Real-time continuous glucose monitoring (CGM) devices help detect glycemic excursions associated with exercise, meals, and insulin dosing in patients with type 1 diabetes (T1D). However, the delay between interstitial and blood glucose may result in CGM underestimating the true change in glycemia during activity. The purpose of this study was to examine CGM discrepancies during exercise and the meal postexercise versus self-monitoring of blood glucose (SMBG). Methods: Seventeen adults with T1D using insulin pump therapy and CGM completed 60 min of aerobic exercise on three occasions. A standardized meal was given 30 min postexercise. SMBG was measured during exercise and in recovery using OmniPod® Personal Diabetes Manager (PDM; Insulet, B