Ontology highlight
ABSTRACT: Purpose
Magnetic resonance imaging of lungs is inherently challenging, but it has become more common with the use of UTE sequences and their relative insensitivity to motion. Spiral UTE sequences have been touted recently as having greater k-space sampling efficiencies than radial UTE, but few are designed for the shorter T2 * of the lung. In this study, FLORET (Fermat looped, orthogonally encoded trajectories), a recently developed spiral 3D-UTE sequence designed for the short T2 * species, was implemented in human lungs for the first time and the images were compared with traditional radial UTE images.Methods
The FLORET sequence was implemented with parameters optimized for lung imaging on healthy and diseased (cystic fibrosis) subjects. On healthy s
SUBMITTER: Willmering MM
PROVIDER: S-EPMC6559861 | biostudies-literature | 2019 Sep
REPOSITORIES: biostudies-literature