Capillary Vibrating Sharp-Edge Spray Ionization (cVSSI) for Voltage-Free Liquid Chromatography-Mass Spectrometry.
Ontology highlight
ABSTRACT: Here, we report a continuous flow-based ionization method, capillary vibrating sharp-edge spray ionization (cVSSI), that nebulizes liquid sample directly at the outlet of a capillary without using high-speed nebulization gas or a high electrical field. cVSSI is built upon the recently reported VSSI principle which nebulizes bulk liquid using vibrating sharp-edges. By attaching a short piece of fused silica capillary on top of the vibrating glass slide in VSSI, liquid is nebulized at the outlet of the capillary as the result of the vibration. Utilizing standard 360-μm OD/100-μm ID capillary, cVSSI works with a wide range of flow rates from 1 μL/min to 1 mL/min. The power consumption is as low as 130 mW. ESI-like MS spectra are obtained for small molecules, peptides, and proteins. Five order