ABSTRACT: Background
Little is known about the ICD performance using enhanced detection algorithms in unselected, non-trial patients. Performance of recent generation ICD equipped with SmartShock™ technology (SST) for detection and conversion of ventricular tachyarrhythmias (VTA) was investigated.Methods
4P was a prospective, multicenter, observational study conducted in 10 Swiss implanting centers. Patients with a Class I indication according to international guidelines were included and received an ICD with SST. ICD discrimination capability was assessed by evaluating SST performance; therapy efficacy was assessed by rate of VTA conversions by ATP and by rescue shocks.Results
Overall, 196 patients were included in the analysis with a mean duration of follow-up of 27.7 month