Optimization of tube voltage in X-ray dark-field chest radiography.
Ontology highlight
ABSTRACT: Grating-based X-ray dark-field imaging is a novel imaging modality which has been refined during the last decade. It exploits the wave-like behaviour of X-radiation and can nowadays be implemented with existing X-ray tubes used in clinical applications. The method is based on the detection of small-angle X-ray scattering, which occurs e.g. at air-tissue-interfaces in the lung or bone-fat interfaces in spongy bone. In contrast to attenuation-based chest X-ray imaging, the optimal tube voltage for dark-field imaging of the thorax has not yet been examined. In this work, dark-field scans with tube voltages ranging from 60 to 120 kVp were performed on a deceased human body. We analyzed the resulting images with respect to subjective and objective image quality, and found that the optimum tube