ABSTRACT: Background
Quantitative MRI can detect early changes in cartilage biochemical components, but its routine clinical implementation is challenging.Purpose
To introduce a novel technique to measure T1 and T2 along radial sections of the hip for accurate and reproducible multiparametric quantitative cartilage assessment in a clinically feasible scan time.Study type
Reproducibility, technical validation.Subjects/phantom
A seven-compartment phantom and three healthy volunteers.Field strength/sequence
A novel MR pulse sequence that simultaneously measures proton density (PD), T1 , and T2 at 3 T was developed. Automatic positioning and semiautomatic cartilage segmentation were implemented to improve consistency and