Establishing and validating of an laboratory information system-based auto-verification system for biochemical test results in cancer patients.


ABSTRACT:

Background

To establish and validate an laboratory information system (LIS)-based auto-verification (AV) system by using large amounts of biochemical test results in cancer patients.

Methods

An algorithm of the AV process was designed for pre-analysis, analysis, and post-analysis. The limit range check was adjusted three times, while the delta check criteria were first replaced by the same patients' historical extremum results. AV rules of 51 biochemical test items were tested by using data of 121 123 samples (6 177 273 tests) in 2016 that were manually reviewed through the simulative i-Vertification software of Roche. The improved and optimal AV rules were programed into our LIS and validated by using 140 113 clinical specimens in 2018.

Results

The AV passing rate fo