Multicenter Laboratory Comparison of Iohexol Measurement.


ABSTRACT:

Background

Iohexol is utilized for measurement of kidney glomerular filtration rate (GFR). Until recently, there have not been available proficiency standards to assist in calibrating a laboratory's results. In view of a shift in calibration at the University of Rochester Medical Center (URMC) laboratory, serving as Central Biochemistry Laboratory for the CKiD study, we performed a multi-centered laboratory comparison.

Methods

Two batches of 30 fortified sera and patient samples from serum or heparinized plasma were sent for duplicate analysis to URMC, University of Minnesota (UMN), Mayo Clinic, and University of Lund. Five proficiency testing materials from Equalis AB were also provided. Iohexol calibration was performed using dilutions of Omnipaque™ 300 and concentrations